Introduction
Contrast is the difference between the luminance of two 
objects. For example, a dark gray ball on a white tablecloth 
has high contrast; if the object were light gray, the contrast 
would be lower
      The definition of contrast can be formalized by the 
equation:
      Micbelson contrast L L L L =? + ( max min max min )/( ) [1]
where Lmax and Lmin are the maximum and minimum
luminance of an image.
      If the background luminance (in our example the 
tablecloth) is constant, Weber’s law can be assumed.
      Weber contrast L L L = ? back back [2]
where L is the luminance of the object and Lback is the 
luminance of the background (or pedestal, in our case the 
tablecloth). Snellen optotypes for visual acuity testing have
generally Weber contrast ≥90% (1).
      The contrast threshold, therefore, is the just noticeable 
difference (jnd) in luminance. Contrast sensitivity (CS), the 
reciprocal of contrast threshold (CS =1/contrast threshold), 
increases as a function of the ability of the observer to 
perceive this difference.
      As it often happens for psychophysical measurement 
units, CS can be expressed as a logarithmic scale to ensure 
differences in signal intensity have the same value across the 
whole spectrum of luminance. 
	The contrast sensitivity function (CSF)
CS can be measured by using grating stimuli. It is well 
known that the size of the stimulus affects CS (2): in a
grating made of dark and light bars, as the thickness of 
the bars decreases (the spatial frequency of the grating 
increases), the amount of contrast required to see the 
grating increases. It is worth recalling that for a serial 
stimulus, the spatial frequency corresponds to the number 
of repetitions per visual space unit (cycles /degree). E.g., a 
1 degree-wide grating made of 10 black bars and 10 white 
bars has a spatial frequency of 10 cycles/deg, where 1 cycle 
is made up of a white bar plus a black bar.
      Luminance change between the stripes is described 
by a square- or sine-wave function. In the first case, the 
luminance of the bars is constant across their extension, in 
the second, luminance is at its maximum along the central 
axis of the white stripes and decreases progressively towards 
the extremes, reaching the minimum along the central axis 
of the dark stripes (Figure 1).
      Contrast threshold is therefore the minimum amount 
of contrast required to detect bars of a given spatial 
frequency. Since CS is the reciprocal of contrast threshold, 
the CSF is a threshold function that describes how contrast 
threshold changes with spatial frequency. It is the result 
of multidimensional analysis since it describes how a 
dependent variable (sensitivity) varies as a function of 
the independent variable (spatial frequency). Conversely, 
the frequency-of-seeing curve, a sigmoid function that 
describes the subject’s response to stimulus intensities, 
is unidimensional and represents the percent of correct 
responses as a function of the variable under examination 
(the strength of the signal).
      The CSF (Figure 2) can be described by a log parabola 
with a peak at the medium frequencies (3–6 c/deg) and 
decay at the high and low spatial frequencies (3,4). 
      In the absence of alterations of the eye and/or the visual 
pathway, the fall-off in sensitivity at high spatial frequencies 
depends on the density of the foveal photoreceptors, 
thereby on the maximum resolution of the visual system; 
indeed, the foveal cone system is characterized by higher 
resolution as well as higher activation threshold than the 
peripheral retina. The decay at low spatial frequencies, 
arguably, depends on lateral inhibition between contiguous 
ganglion cells (the interaction between the opposing signals 
of center and surround of neighboring detectors leads to 
spatial attenuation of slow luminance gradients). Therefore, 
with the CSF, it is possible to estimate the visual acuity, 
which corresponds to the cut-off frequency represented 
by the point at which the function intercepts the abscissa, 
namely the finest gratings that can be detected at 100% 
contrast.
	P-mediated and M-mediated CSF
The anatomofunctional organization of the visual system 
relies on two cellular pathways: the magnocellular and 
the parvocellular system (M- and P-system, respectively).
The M-system is responsible for the analysis of global and 
rapidly moving configurations: in effect, it is particularly 
sensitive to stimuli made of components with low spatial 
and high temporal frequencies, even more, if at low 
luminance levels (5,6). On the other hand, it is less activated 
by high spatial frequencies and low temporal frequencies (7). 
As an example, consider large clear and dark bars that 
quickly translate or invert polarity, generating a flicker. 
Damage to the M-lateral geniculate cells determines 
reduced CS for stimuli with low spatial frequency 
(1 cycle/degree) and high temporal frequencies (10 Hz 
or higher) (7). The P-system is sensitive to static details, 
therefore it is responsive to high spatial frequencies and low 
temporal frequencies, particularly at high luminance levels 
(e.g., a stationary grating made up of narrow dark and light 
stripes).
      The CS curve shown in Figure 2 and Figure 3 is the result of 
the complementary activity of the M- and P-system. The cutoff or transition point from M/P activity is located between 0.2 
and 3.5 cycles/degree (8) or at about 1.5 cycles/degree (9).
      Even though the detection of a sine-wave grating at low 
spatial frequency is carried out by the M-cells, the P-system 
is globally more responsive, so its activation extends to a 
certain degree to the M-domain. Maunsell and colleagues, 
for example, found a moderate parvocellular activation in 
the middle temporal visual area, responsible for processing 
M-mediated motion perception (10). For this reason, an 
impairment of the P-system decreases CS not only at the 
high spatial frequencies, as expected, but to some extent 
even at the low frequencies [1 cycle/degree (11,12); 2 cycles/
degree (13)]. In fact, Merigan and Eskin observed a deficit 
of CS for stationary gratings up to 0.5 cycles/degree after 
selective destruction of the geniculate parvocellular layers 
in monkeys. However, the deficit was less evident when the 
temporal frequency was increased, as the result of a greater 
M-activation (14). The response is saturated at spatial 
frequencies of about 1 cycle/degree and temporal frequency 
of 10 Hz: spatial frequencies of less than 1 cycle/degree with a 
temporal frequency of at least 10 Hz, at low luminance levels, 
can be considered, with due caution, the best experimental 
parameters to estimate the magnocellular activity.
	Types of CS loss in patients
By examining the CSF, four patterns of alteration can be 
identified (15): 
(I) A defect limited to the high spatial frequencies that 
reflects a reduction of visual acuity as commonly 
measured with high-contrast symbols; this defect
is typical of uncorrected ametropia or amblyopia 
(16,17); 
(II) A defect that encompasses all the frequencies. In 
this case, low CS at the higher spatial frequencies 
is responsible for reduced visual acuity and is 
associated with a deficit at medium and low spatial 
frequencies. This is typical of lenticular and neural 
aging (18,19);
(III) A defect limited to the mid-range frequencies as in 
the case of multiple sclerosis (20);
(IV) A defect to the medium-low spatial frequencies 
in case of glaucoma, optic neuritis, papilledema, 
multiple sclerosis, or diabetes (21,22), or at the 
low spatial frequencies in dyslexia, as postulated 
by the so-called magnocellular theory. A class of 
dyslexics struggle to read because of a deficit in the 
dorsal magnocellular visual pathway [see (23) for a 
review on this topic]. In line with this, some studies 
found reduced CS at low spatial frequencies in 
these patients (24,25), especially at high temporal 
frequencies and meso-scotopic conditions, i.e., 
those mediated by the magnocellular system (26). 
Nevertheless, the finding is arguable, as the 
parvocellular system contributes significantly to 
contrast threshold at all the frequencies (27).
To plot accurately the threshold function, the entire 
bandwidth should be evaluated: yet, it is time-consuming, 
especially within the clinical setting. To overcome this 
problem and quickly categorize the CS loss, Cobo-Lewis (28) 
proposed an adaptive Bayesian procedure based on the socalled Minimum Estimated Expected Entropy (MEEE).
     To establish if the visual function is normal, estimating as 
few as two functional parameters is stated to be sufficient, 
namely CS at high spatial frequencies, that reflects visual 
acuity, and at mid-range spatial frequencies (29). The 
combination of these two measurements, indeed, provides a 
good estimate of the CSF with a reduced number of trials.
	Clinical assessment of CS
      The first rudimentary test to investigate CS was 
introduced in the 18th century by Boguer. It consisted of 
two candles, one placed near a screen and the other farther 
away. An opaque bar positioned between the two candles 
at a variable distance cast a shadow on the screen. The 
amount of contrast was given by the differential luminance 
between the background (the screen) and the shadow (the 
target). The bar was moved away until the patient was 
no longer able to detect the shadow. In the 19th century, 
Bjerrum introduced the first low contrast optotype 
(alphanumeric stimuli with a contrast of 9%, 20%, 30%, 
and 40%). Later on, in the 1950s, Fortuin adopted 
optotypes with various levels of luminance. Finally, in 
the 1960s, sine-wave gratings started to be used (4) [for a 
historical review of the tests see (30)].
      Since then, more rigorous and sophisticated exams for 
the measurement of CS have been developed, like optotypic 
tables or psychophysical algorithms implemented by 
computerized techniques.
	The Pelli-Robson CS-chart
      The Pelli-Robson chart (29) depicts lines of letters invariant 
in size but changing in contrast, expressed as Weber’s 
fraction (Figure 4). According to the authors, this solution is 
preferable, since the recognition of letters is a more familiar 
task than the detection of gratings. This table is calibrated 
for a distance of 3 meters and is made up of lines of letters 
0.5 degrees wide (Sloan font) arranged one under the other. 
Presenting the table at 1 meter is especially suitable to 
test low-vision patients: at this viewing distance the letters 
subtend 2.8 degrees of visual angle. 
      Each line consists of two triplets of letters: The contrast of 
the right triplet is lower than the left one by a factor of 1/√2 
(i.e., 0.15 log units). Contrast is reduced by the same amount 
not only between the two triplets of the line but also across 
the lines. The observer is asked to read each line from the top 
to the bottom of the chart. Contrast threshold is computed 
as the amount of contrast of the triplet preceding the one 
in which two letters have not been recognized. As for visual 
acuity, the exam does not include the presentation of all the 
stimuli on the table but ends when the observer is no longer 
able to recognize two letters of the triplet
      Pelli estimated that letters 0.5 deg in size are suitable for 
measuring CS at spatial frequencies between 3 and 5 cycles/
degree, i.e., at the medium frequencies around the middle 
of the CSF, which the human visual system is optimally 
sensitive to. He pinpointed that this bandwidth provides as 
much information on CS as would be clinically useful. For 
this reason, in association with the measurement of visual 
acuity, the Pelli-Robson chart is enough to characterize the 
global visual function of the subject. The target probability 
for threshold estimation depends on the response model. 
Assuming it is 26-AFC (i.e., as many alternatives as the 
letters of the alphabet, even if, unknown to the observer, 
only 10 characters for the Sloan font are presented), 
implicit version [for a definition of standard- and implicitAFC version see (31)], the guess rate is 3.8%. Therefore, 
a target probability higher than (100% + 3.8%)/2 =51.9% 
is required. The recognition of 2 out of three characters of 
the triplet means a 66% proportion of correct responses. 
A procedure similar to the Pelli-Robson is the Mars Letter 
Contrast Sensitivity Test. The Mars Letter Contrast 
Sensitivity Test (32) is made of a set of three alphanumeric 
charts (one for the right eye, one for the left eye, and one 
for binocular vision) calibrated for near vision distance. 
Like in the Pelli-Robson chart, the size of the symbols is 
constant. In each table, the contrast level is progressively 
reduced letter-by-letter by 0.04 log unit steps (48 contrast 
levels) and the score is computed from the letter (or 
number) with the smallest amount of contrast that can be 
perceived in each table. 
	The Low-Contrast Sloan Letter Charts (LCSLCs)
The Pelli-Robson charts measure CS of letters with a 
predetermined size. In turn, the LCSLCs (33) estimate 
visual acuity at a certain amount of contrast. Seven tables 
(illuminance: 861-1076 lux) with lines of Sloan letters 
in ETDRS-like format are administered at a distance of 
2 meters. The seven tables differ in contrast, from 100% to 
0.6% (Figure 5). The subject is asked to read the letters in 
each table starting from the upper line, then an acuity score 
corresponding to the number of letters identified correctly
is computed for each contrast level (the scoring system 
is similar to the letter-by-letter ETDRS method for the 
assessment of visual acuity). 
      As the contrast of the table is lower, the recognition 
task becomes more and more demanding. Therefore, 
the letter score for each chart is inversely correlated to 
the visual threshold at that contrast value: the higher the 
score, the higher the sensitivity at that contrast level. The 
psychophysical procedure resembles that of Pelli-Robson 
(26-AFC response model, implicit variant, method of 
constant stimuli) but differs since, as explained, the LCSLCs 
measure visual acuity as a function of a given amount of 
contrast: so, the LCSLCs do not directly estimate CS, but 
the effect of contrast on visual acuity.
	The Arden gratings (grating CS test)
The Arden test (34) makes use of 6 panels with luminance 
130–150 cd·m-2, each reproducing a sine-wave grating 
with different spatial frequencies (0.2, 0.4, 0.8, 1.6, 3.2, 
6.4 c/deg). The contrast of the gratings increases across the 
panel from top to bottom by a value equal to 0.088 log units 
(1.6 dB) every 1.1 cm (Figure 6). The observer is given the 
first panel at a distance of 50 cm, completely covered with a 
sheet of paper except for the upper portion, where the sinewave bars are not detectable. The direction of presentation 
(from minimum to maximum contrast) is therefore opposite 
to that adopted in the Pelli-Robson chart (from maximum 
to minimum contrast letters). The plate is then slowly 
uncovered. The observer is asked to report the position 
of the sheet as soon as he can detect the bars. A scale on 
the side of the plate converts the position of the sheet into 
the threshold. Then, the second plate is presented and 
the examination is iterated for all the plates so to obtain 
a threshold per spatial frequency tested. The thresholds 
obtained at the different spatial frequencies allow plotting 
the CSF.
      The procedure follows the method of limits for 
continuous stimulation (adjustment)
	The 4AFC CS test
Vaegan
and Halliday (35) proposed a 4-AFC response 
model to assess CS. Parts of Arden’s gratings are cut into a 
series of discs that differ in contrast and spatial frequency 
to be presented in four different orientations (horizontal, 
vertical, left, or right oblique). The same six frequencies of 
the Arden gratings are administered. The disk is presented 
with a certain orientation and the observer is forced to 
report the orientation of the grating. The threshold 
corresponds to the grating with the lowest contrast whose 
orientation is recognized at each spatial frequency.
	The Cambridge low contrast gratings
The Cambridge low contrast gratings (36) administer 
square-wave stimuli to test CS within a range of eleven 
values, from 0.89 to 2.85 log contrast (i.e., from 13% to 
0.11% contrast value), at a spatial frequency of 4 cycles/deg. 
The authors decided to test only the spatial frequency of 
4 c/deg since they assumed that an alteration of any other 
frequencies always implies a deficit at 4 c/deg. Moreover, 
the peak in CS is observed at spatial frequencies between 3 
and 6 c/deg (37): these are the spatial frequencies that are 
necessary to perform common visual tasks (38). A grating 
and a null (uniform) stimulus are presented at a distance 
of 6 meters; the observer is asked to indicate the target 
(the grating). The test is repeated four times at each level 
of contrast: In the original version, a score is obtained 
from the total number of detected gratings. The score 
is then converted into CS. The procedure uses a forcedchoice response design (2AFC), method of constant stimuli 
(4 constant stimuli for each contrast value), detection task. 
Jones and colleagues found that the repeatability varies by 
about one-third of the total performance range, therefore 
they recommend caution in monitoring CS with this 
method (39).
	The Sine Wave Contrast Test Vistech CS charts (Ginsburg’s 
gratings)
Ginsburg (40) developed a test based on the detection 
of sine-wave gratings vertically oriented or rotated by 
15 degrees clockwise or counterclockwise. Five different 
spatial frequencies are administered (1.5, 3, 6, 12, and 
18 c/deg) at 9 contrast levels (non-constant step size with an 
average value of 0.25 logarithmic units: Figure 7). There is 
a version for far distance (3 meters) and a version for near 
distance (40 centimeters). 
	The method of adjustment with manual contrast control
Another technique developed by Vaegan and Halliday (35) 
makes use of the method of adjustment and administers 
gratings with predefined spatial frequency. The contrast 
of each grating is progressively increased by the observer 
until the pattern is detected (the operator ensures during 
the examination that the speed at which the observer 
                             
increases the contrast of the grating is constant and around 
0.5 dB/sec). This method is quick and simple but is strictly 
dependent on the observer’s response criterion. According 
to the authors, decreasing series are not suitable since 
they are much more variable. In effect, gradual reduction 
of intensities leads the observer to adapt to the previous 
high contrast stimulation, resulting in an overestimation 
of the threshold (1). Another method of adjustment, called 
von Békésy tracking, administers ascending and descending 
series of contrast levels; after a predetermined number of 
reversals in the trend of the responses, detection threshold 
is estimated as the mean of the reversal points. However, 
repeatability appeared to be lower than for the classical 
method of adjustment with increasing series (41).
	The Freiburg visual Acuity and Contrast Test (FrACT)
FrACT
(42,43) assesses not only high-contrast visual acuity
but also CS for a given angular dimension of the stimulus. 
As for the measurement of visual acuity, it is an 8-AFC 
response design, with the detection task focused on the 
orientation of Landolt’s “C”. Presentations are guided by 
the best PEST psychophysical procedure. Best PEST is 
a Bayesian procedure (like the MEEE method) than can 
quickly estimate the threshold, assuming that the slope of 
the psychometric function is known (31).
	The Holladay Automated Contrast Sensitivity Testing 
System (HACSS)
The HACSS (44) administers concentric circular sine-wave 
stimuli with a spatial frequency of 1.5, 3, 6, 12, 18 c/deg. 
At each frequency, the stimulus is presented starting from 
a contrast level of 50% (viewing distance: 4 meters). The 
observer must report if he detects a concentric target 
(Figure 8) or if he sees a uniform circle. The operator 
presses a key corresponding to the patient’s response. The 
program includes a protocol for photopic vision (luminance: 
85 cd·m-2) and one for mesopic vision (luminance: 3 cd·m-2) 
by applying a filter to the monitor
Contrast is initially reduced by 0.3 log units for each 
correct answer, then by 0.1 log units near the threshold. In the case of a wrong response, the contrast is increased by 
0.3 log units, which becomes 0.2 after the second wrong 
response. The threshold corresponds to the lowest contrast 
at which the subject detects two out of three stimuli. The 
threshold is computed as the mean of the two best results 
before the next incorrect answer. The CSF is obtained 
by repeating the procedure for each spatial frequency. 
During the exam, null stimuli are presented as false-positive 
controls and a reliability index is computed.
      HACSS follows a simple up-down staircase procedure 
associated with a yes/no response model, but the step size is 
reduced at a certain point of the examination.
	Which is the best way to measure CS?
It should be borne in mind that the estimation of CS, in 
terms of absolute values and shape of the CSF, depends on 
the response model employed and on the psychophysical 
technique (45). Alternative forced-choice (AFC) responses 
should be preferred over criterion-dependent methods (yes/
no model) because they minimize the undesired effect of the 
subjective criterion, increasing reliability and repeatability 
(18,35). Within the framework of the signal detection theory, 
the subjective criterion varies depending on the degree 
of confidence of the subject in detecting the signal (46). 
Contrary to the yes/no response model, alternative-forced 
choice designs rule out this problem. Pelli and colleagues 
suggested using at least 4 or 5 alternatives to minimize the 
guessing rate and make CS assessment more accurate (29).
      Which is the best stimulus to be administered is a 
debated topic. Letter charts measure recognition while 
gratings refer to a detection task (31,47): the two thresholds 
do not match and, as pointed out by Leguire, some clinical 
conditions could influence them to a different extent (47). 
The same author suggested that recognition of letters 
is influenced by their different legibility, even if Pelli 
and Regan considered this aspect not relevant (48,49). 
What is substantial is that the Pelli-Robson chart does 
not provide the shape of the CSF, but measures CS at a 
predetermined spatial frequency (that is a function of the 
adopted viewing distance). It follows that it is not suitable 
to detect high frequency (type 1) CS losses (47,50). Despite 
this shortcoming, letter charts are simple, quick and easy to 
be administered, and provide reliable results. Moreover, the 
Pelli-Robson chart does not suffer from ceiling and floor 
effects, and reliability and repeatability are better compared 
to gratings (29,51,52) and to the computerized test FrACT, 
that makes use of Landolt Cs (53).
      Low-contrast optotypes (LCSLC) have a smaller step 
size and use a higher number of forced-choice alternatives 
compared to grating-based CS tests like the Vistech chart 
(0.10 vs. 0.25 log steps and 26 vs. 3 AFC, respectively), 
resulting in higher precision and reliability compared to 
the former (33,51). Their peculiarity is that they assess 
how visual acuity is affected by contrast, i.e., they measure 
the smallest letter that can be resolved at a predetermined 
contrast, but do not provide a direct estimate of CS.
Indeed, test-retest reliability and validity of grating tests 
are reported to be quite poor [in particular in the original 
version of the Vistech (51,52)], especially at the lower 
spatial frequencies. Despite they are suitable for screening 
purposes (34), this suggests caution in clinical use (52). It 
remains that gratings show a substantial advantage over 
letters and other optotypes, that is they plot the CSF: a 
valuable piece of information about the overall integrity 
of the visual system (4). Sinusoidal gratings should be 
preferred to square waves because of their luminance 
profile that is robust against defocus, optical aberrations, 
diffraction, or light scatter (1). 
      Regarding the psychophysical procedure, non-adaptive 
procedures like the method of limits with increasing 
(ascendent) stimuli provides better reliability than the 
method of adjustment, and its variant von Békési tracking (41). 
The method of constant stimuli, indeed, is the most 
simple way to compute the psychometric function of the 
observer (31) but is too time-consuming, i.e., not efficient, 
for the clinical purpose (54). 
       Adaptive strategies reduce examination time. Parametric 
procedures like best PEST (adopted by the FrACT) are 
accurate and less time-consuming than the nonparametric 
counterparts (31), while test-retest variability is quite 
similar (55). Of note, patients may have difficulty 
familiarizing themselves with parametric strategies, due to 
the fast convergence toward the threshold since the initial 
phase of the examination (55). 
      A final question arises about whether different visual 
pathways can be investigated via different techniques. As 
highlighted by an anonymous reviewer, it is worth recalling 
that all the tests discussed so far measure CS at an overall 
perceptual level. Even if some of them are focused on a 
specific range of spatial frequencies, none of them selectively 
target, i.e., isolate, the magnocellular or the parvocellular 
CSF; due to the wide overlapping of the P- and M-function, 
indeed, both are always involved in the CS estimates. 
	The critical fusion frequency (CFF)
The investigations reported up to this point addressed 
the problem from the perspective of spatial contrast, and 
analyzed CS as a function of the visual angle subtended by 
the target in stationary temporal conditions. A different 
strand of research focused on temporal contrast, that is to 
say on the sensitivity of the visual system as a function of the 
flicker frequency of a grating with fixed spatial frequency
       Temporal CS in the absence of a spatial component is 
defined as the sensitivity to luminance differences within 
the temporal domain, i.e., as a function of time. Likewise, 
the limit to temporal sensitivity of the visual system is 
reflected in the CFF that is the temporal frequency at which 
a flashing target is no longer perceivable as flickering.
      This threshold depends on the neuronal limitation in
encoding signals in the temporal domain: the neurons are 
unable to modulate tachistoscopic information beyond a 
certain temporal frequency: it follows that the flickering 
stimulus is perceived as stationary.
      CFF is considered a good indicator of the integrity of 
visual temporal processing, abnormal in encephalopathies, 
ocular conditions such as multiple sclerosis with 
involvement of the optical pathways (56,57), macular 
degeneration (58), cataract (59), and glaucoma (60).
      The amount of temporal modulation of the visual 
stimulus is the modulation amplitude, that is the oscillations 
of luminance over time. Modulation amplitude can be 
expressed as a modulation percentage (Figure 9).
      The performance of the visual system as a function of 
the temporal frequency and the modulation percentage is 
described by the temporal modulation transfer function 
(TMTF). TMTF implies a multidimensional analysis, like 
the CSF, and its shape is similar to the latter, with a central 
peak, a steep decay at high temporal frequencies, and a more 
gradual decrease at low temporal frequencies (Figure 10). 
According to this paradigm, the CFF measures the visual 
temporal resolution. It corresponds to the temporal 
frequency at which the flicker is no longer detected (i.e., the 
threshold) at a certain modulation percentage.
CFF increases with the logarithm of the size of the 
target: the greater the stimulus, the greater the CFF,
thereby the easier the detection of the flicker (61). This 
trend is in agreement with the different retinal distribution 
of parvo- and magnocells: the magnocellular (or transient) 
system is more represented in the peripheral retina, 
therefore its activation grows as the target centered to 
the fovea is made larger. CFF varies also as a function of 
contrast, adaptation to light, aging, and, above all, as a 
function of stimulus intensity (62,63): in fact, CFF increases 
directly with the logarithm of the target’s luminance. The 
relationship is expressed by the Ferry-Porter law:
where I is the stimulus luminance, I0 is the threshold 
luminance, and k has a typical value of 10–30 Hz (64). 
      To date, it is not simple to establish which is the best 
paradigm for estimating the CFF. Eisen-Enosh and 
colleagues compared the method of limits associated with a 
yes/no response model, the method of constant stimuli, and 
the staircase both coupled to a 2AFC temporal paradigm. 
Test-retest reliability is satisfactory for all three methods, 
albeit slightly lower for the limits (65). Since the staircase 
is an adaptive procedure that combines good accuracy 
with a short execution time, it may be considered the most 
suitable.
	Conclusions
The visual system relies on multiple channels to transmit 
information from the retina to the higher centers.
In the spatial domain, high contrast stimuli, used 
for measuring central visual acuity, recruit mainly the 
parvocellular pathway. As a consequence, estimating visual 
acuity may not yield a comprehensive overview of the 
functional status of the visual system: as a matter of fact, 
there are pathological conditions that affect preferentially
the magnocellular system. In these cases, a decline in CS, 
both in the spatial and temporal domain, takes place in the 
face of a normal capacity to discriminate and identify fine 
details. The importance of measuring CS relies upon this 
basis, and is demonstrated by the considerable number 
of psychophysical procedures proposed so far for its 
assessment as a complement to visual acuity. It is likely that, 
among those reported in this paper, there is not a technique 
that should be preferred to the others, since each of them 
has its pros and cons, some are more suitable for screening 
purposes, others are more suitable for evaluating the effect 
of this function on specific tasks like, for example, reading. 
      Future directions in this field involve virtual realitybased tests (66) or the use of novel stimuli that make 
use of illusory motion to generate contrast, and whose 
effectiveness in measuring CS in patients with retinal vein 
occlusion and age-related macular degeneration has been 
recently reported (67,68). Finally, a method to measure CS
based on real-time detection of the optokinetic response 
has been proposed and found effective in detecting the 
effect of defocus on CS in emmetropic subjects (69). The 
development of these new techniques will substantially 
contribute to improve the early diagnosis and treatment of 
ophthalmic pathologies based on CS.
	Acknowledgments
Funding: None.